Wireless communicative stent for follow-up of abdominal aortic aneurysm,
Destrade, Michel
Destrade, Michel
Loading...
Repository DOI
Publication Date
2006
Type
Conference Paper
Downloads
Citation
J. Mazeyrat, O. Romain, P. Garda, E. Flecher, M. Karouia, P. Leprince, P.-Y. Lagree, M. Destrade (2006) Wireless communicative stent for follow-up of abdominal aortic aneurysm . In: IEEE eds. 2006 IEEE Biomedical Circuits and Systems Conference (IEEE-BioCAS 2006) London, 29/11/2006- 01/12/2006
Abstract
An abdominal aortic aneurysm (AAA) is a dilatation of the aorta at the abdominal level, which rupture is a life threatening complication. Recent treatment of AAA consists in endovascular treatment with covered stent grafts. Despite improving devices, this treatment is still associated with close to 25% of failure related to persisting pressure into the excluded aneurismal sac. The follow-up becomes thus crucial and demands frequent examinations (CT-scan, IRM) which are not so liable given the complications. In order to evaluate the post-operative period of an AAA treatment, we designed a communicative stent, comprising of an integrated pressure sensor. This paper presents the conception of a communicative sensor, the elaboration of a numerical model, and the development of an experimental testbench constituting the aortic flux across an AAA and allowing the optimization and validation of the measurement principle.
Funder
Publisher
IEEE
Publisher DOI
10.1109/BIOCAS.2006.4600352
Rights